The application of an ultraviolet (UV) light-emitting diode (LED) to on-line sample concentration/fluorescence detection in capillary electrophoresis (CE) is described. The utility of a UV-LED (peak emission wavelength at 380 nm, ≃2 mW) for fluorescence detection was demonstrated by examining both a naturally fluorescent (riboflavin) compound and a non-fluorescent compound (tryptophan), respectively. The detection limit for riboflavin was determined to be 0.2 ppm by the normal MEKC mode, which was improved to 3 - 7 ppb when dynamic pH-junction technique was applied. On the other hand, the detection limit of the tryptophan derivative was determined to be 1.5 ppm using the MEKC mode, which was improved to 3 ppb when the sweeping-MEKC mode was applied. In an analysis of an actual sample, the concentrations of riboflavin in beer, and tryptophan in urine and milk samples were determined, respectively. 2006
| 原文 | 英語 |
|---|---|
| 頁(從 - 到) | 235-240 |
| 頁數 | 6 |
| 期刊 | Analytical Sciences |
| 卷 | 22 |
| 發行號 | 2 |
| DOIs | |
| 出版狀態 | 已發佈 - 2006 2月 |