Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom

H. L. Chao, W. L. Chen, C. C. Hu, J. K. Wu, C. J. Wu, J. C.H. Cheng*

*此作品的通信作者

研究成果: 雜誌貢獻期刊論文同行評審

2 引文 斯高帕斯(Scopus)

摘要

Purpose: The accurate volumetric calculation of moving targets/organs is required to use cone-beam computed tomography (CBCT) for replanning purposes. This study was aimed to correct the reconstructed volume losses of moving phantoms by phase-specific CBCT. Materials and methods: Planning fan-beam CT (FBCT) of five hepatobiliary/gastrointestinal/pancreatic cancer patients were acquired under active breathing control and compared with free-breathing CBCT for kidney volumes. Three different-sized ball phantoms were scanned by FBCT and CBCT. Images were imported to a planning system to compare the reconstructed volumes. The phantoms were moved longitudinally on an oscillator with different amplitudes/frequencies. The phase-specific projections of CBCT for moving phantoms were selected for volume reconstruction. Results: The differences in reconstructed volumes of static small, medium, large phantoms between FBCT and CBCT were -∈6.7%, -∈2.3%, and -∈2.0%, respectively. With amplitudes of 7.5-20 mm and frequencies of 8-16 oscillations/min, volume losses on CBCT were comparable with FBCT in large moving phantoms (range 9.1-27.2%). Amplitudes were more subject to volume losses than frequencies. On phase-specific CBCT, volume losses were reduced to 2.3-6.5% by reconstruction using 2-3 projections at end/midoscillation phase. Conclusion: Amplitude had more impact than frequency on volume losses of moving phantoms on CBCT. Phase-specific CBCT reduced volume losses.

原文英語
頁(從 - 到)77-83
頁數7
期刊Strahlentherapie und Onkologie
188
發行號1
DOIs
出版狀態已發佈 - 2012 1月

ASJC Scopus subject areas

  • 放射學、核子醫學和影像學
  • 腫瘤科

指紋

深入研究「Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom」主題。共同形成了獨特的指紋。

引用此