Measurements of natural uranium concentration and isotopic composition with permil-level precision by inductively coupled plasma-quadrupole mass spectrometry

Chuan Chou Shen, Huei Ting Lin, Mei Fei Chu, Ein Fen Yu, Xianfeng Wang, Jeffrey A. Dorale

研究成果: 雜誌貢獻文章

10 引文 斯高帕斯(Scopus)

摘要

A new analytical technique using inductively coupled plasma-quadrupole mass spectrometry (ICPQMS) has been developed that produces permil-level precision in the measurement of uranium concentration ([U]) and isotopic composition (δ234U) in natural materials. A 233U-236U double spike method was used to correct for mass fractionation during analysis. To correct for ratio drifting, samples were bracketed by uranium standard measurements. A sensitivity of 6-7 × 108 cps/ppm was generated with a sample solution uptake rate of 30 mL/min. With a measurement time of 15-20 min, standards of 30-ng uranium produced a within-run precision better than 3%(±2 R.S.D.) for δ234U and better than 2%for [U]. Replicate measurements made on standards show that a between-run reproducibility of 3.5%for δ234U and 2% for [U] can be achieved. ICP-QMS data of δ234U and [U] in seawater, coral, and speleothem materials are consistent with the data measured by other ICP-MS and TIMS techniques. Advantages of the ICP-QMS method include low cost, easy maintenance, simple instrumental operation, and few sample preparation steps. Sample size requirements are small, such as 10-14 mg of coral material. The results demonstrate that this technique can be applied to natural samples with various matrices.

原文英語
文章編號Q09005
期刊Geochemistry, Geophysics, Geosystems
7
發行號9
DOIs
出版狀態已發佈 - 2006 九月 1

    指紋

ASJC Scopus subject areas

  • Geophysics
  • Geochemistry and Petrology

引用此