Magnetocardiography of animals in magnetically shielded environment with active compensation

Herng-Er Horng*, S. H. Liao, S. J. Hsu, Hong-Chang Yang, J. Y. Wu, C. C. Chen, C. H. Wu, C. C. Wu

*此作品的通信作者

研究成果: 雜誌貢獻期刊論文同行評審

1 引文 斯高帕斯(Scopus)

摘要

A high-Tc 1st-order electronic superconducting quantum interference device (SQUID) gradiometer system is constructed to study the magnetocardiogram (MCG) of rabbits in a moderately magnetically shielded environment with active compensation. In the noisy hospital environment, the noise cannot be completely reduced with the 1st-order gradiometer, therefore, a reference SQUID with active compensation was used to further reduce the noise level leaking into the room. The MCG system was equipped with a x-y translation bed. We used a low-pass filter with the cut off frequency at 44 Hz, a high-pass filter with the cut off frequency at 0.1 Hz and the 60 Hz notch filter to reduce the power line interference. The noise level of the 1st order gradiometer MCG system in this moderately magnetically shielded room was about 1 pT/Hz1/2 at 1 Hz. The MCG of a normal rabbits was measured with this system and a MCG contour map and a current density distribution was constructed.