Direct determination of trace metals at extremely low levels in high purity water is described. The method incorporates an in situ preconcentration technique in the graphite furnace of the atomic absorption spectrometer. The in situ preconcentration process involves multiple injections and evaporation in apyrocoated graphite tube. The feasibility of the proposed method of multiple injections (up to 9×90 μl) prior to analysis has been investigated. A SLRS-2 certified water sample has been used for quality control, and the results are in good agreement with the certified value and with determination by standard addition method. Typically, up to 0.8 ml of sample could be analyzed. Due to the relatively large sample volume, the detection limits for the elements of interest are in the range of 3-9 ng/l.
原文 | 英語 |
---|---|
頁(從 - 到) | 117-123 |
頁數 | 7 |
期刊 | Journal of Radioanalytical and Nuclear Chemistry Articles |
卷 | 172 |
發行號 | 1 |
DOIs | |
出版狀態 | 已發佈 - 1993 8月 |
對外發佈 | 是 |
ASJC Scopus subject areas
- 分析化學
- 核能與工程
- 放射學、核子醫學和影像學
- 污染
- 光譜
- 公共衛生、環境和職業健康
- 健康、毒理學和誘變