In situ preconcentration of trace metals in high purity water onto graphite tube by multiple injections followed by graphite furnace atomic absorption spectrometry

Nae Wen Kuo*, Chi Ren Lan, Z. B. Alfassi

*此作品的通信作者

研究成果: 雜誌貢獻期刊論文同行評審

5 引文 斯高帕斯(Scopus)

摘要

Direct determination of trace metals at extremely low levels in high purity water is described. The method incorporates an in situ preconcentration technique in the graphite furnace of the atomic absorption spectrometer. The in situ preconcentration process involves multiple injections and evaporation in apyrocoated graphite tube. The feasibility of the proposed method of multiple injections (up to 9×90 μl) prior to analysis has been investigated. A SLRS-2 certified water sample has been used for quality control, and the results are in good agreement with the certified value and with determination by standard addition method. Typically, up to 0.8 ml of sample could be analyzed. Due to the relatively large sample volume, the detection limits for the elements of interest are in the range of 3-9 ng/l.

原文英語
頁(從 - 到)117-123
頁數7
期刊Journal of Radioanalytical and Nuclear Chemistry Articles
172
發行號1
DOIs
出版狀態已發佈 - 1993 8月
對外發佈

ASJC Scopus subject areas

  • 分析化學
  • 核能與工程
  • 放射學、核子醫學和影像學
  • 污染
  • 光譜
  • 公共衛生、環境和職業健康
  • 健康、毒理學和誘變

指紋

深入研究「In situ preconcentration of trace metals in high purity water onto graphite tube by multiple injections followed by graphite furnace atomic absorption spectrometry」主題。共同形成了獨特的指紋。

引用此