Implantable stimulator for epileptic seizure suppression with loading impedance adaptability

Chun Yu Lin, Wei Ling Chen, Ming Dou Ker

研究成果: 雜誌貢獻期刊論文同行評審

26 引文 斯高帕斯(Scopus)

摘要

The implantable stimulator for epileptic seizure suppression with loading impedance adaptability was proposed in this work. The stimulator consisted of the high voltage generator, output driver, adaptor, and switches, can constantly provide the required 40-μ A stimulus currents, as the loading impedance varied within 10 kΩ-300 kΩ. The performances of this design have been successfully verified in silicon chip fabricated by a 0.35-μm 3.3-V/24-V CMOS process. The power consumption of this work was only 1.1 mW-1.4 mW. The animal test results with the fabricated chip of proposed design have successfully verified in the Long-Evans rats with epileptic seizures.

原文英語
文章編號6231703
頁(從 - 到)196-203
頁數8
期刊IEEE Transactions on Biomedical Circuits and Systems
7
發行號2
DOIs
出版狀態已發佈 - 2013
對外發佈

ASJC Scopus subject areas

  • 生物醫學工程
  • 電氣與電子工程

指紋

深入研究「Implantable stimulator for epileptic seizure suppression with loading impedance adaptability」主題。共同形成了獨特的指紋。