High-Tc SQUID magnetocardiography imaging system

Hong Chang Yang, S. Y. Hung, C. H. Wu, J. C. Chen, S. J. Hsu, S. H. Liao, H. E. Horng

    研究成果: 雜誌貢獻文章同行評審

    摘要

    We set up a high-Tc SQUID system for magnetocardiography (MCG) in a moderately magnetically shielded room. The electronically balanced gradiometer consists of superconducting quantum interference device (SQUID) magnetometer. One reference SQUID was mounted above the sensing SQUID while the sensing SQUID is seated at the bottom of the cryostat. The baseline of the gradiometer is varied from 5 cm to 7 cm. The output of the MCG signal was filtered with the band pass filter (0.5 - 40 Hz) and the power-line filter. The MCG system was used to detect the magnetic signal of the human heart. Equivalent current sources were used to study the inverse problem.

    原文英語
    期刊Neurology and Clinical Neurophysiology
    2004
    出版狀態已發佈 - 2004 一月 1

    ASJC Scopus subject areas

    • Neuroscience(all)

    指紋 深入研究「High-T<sub>c</sub> SQUID magnetocardiography imaging system」主題。共同形成了獨特的指紋。

    引用此