摘要
Field-dependent NMR studies of bio-molecular systems using a sample shuttling hardware operating on a high-field NMR apparatus have provided valuable structural and dynamic information. We have recently published a design of a compact sample transportation device, called “field-cycler”, which was installed in a commercial spectrometer and which provided highly precise positioning and stability during high speed shuttling. In this communication, we demonstrate the first use of a sample shuttling device on a commercial high field standard bore NMR spectrometer, equipped with a commercial triple resonance cryogenically cooled NMR probe. The performance and robustness of the hardware operating in 1D and 2D field cycling experiments, as well as the impact of the sample shuttling time on the signal intensity are discussed.
原文 | 英語 |
---|---|
頁(從 - 到) | 187-194 |
頁數 | 8 |
期刊 | Journal of Biomolecular NMR |
卷 | 66 |
發行號 | 3 |
DOIs | |
出版狀態 | 已發佈 - 2016 11月 1 |
ASJC Scopus subject areas
- 生物化學