Effect of prolonged pressure equalization on final drifting during pressure wire studies

Chien Boon Jong*, Tsui Shan Lu, Lin Lin, Tsung Yan Chen, Min Tsun Liao, Jui Cheng Kuo

*此作品的通信作者

研究成果: 雜誌貢獻期刊論文同行評審

摘要

Pressure drifting is a troublesome error in invasive coronary function tests. This study aimed to evaluate the relationship between prolonged and short-time pressure equalizations in pressure drifting. Pressure drifting was defined as the pressure gradient between the mean pressure of the distal wire sensor (Pd) and aortic pressure (Pa) when the wire was withdrawn to the tip of the guiding catheter. Significant drifts 1 and 2 were defined as the absolute values of pressure gradients > 2 and > 3 mmHg, respectively. A logistic regression model was used to evaluate the associations between prolonged pressure equalization and each pressure drifting. The prolonged pressure equalization strategy was associated with a lower incidence of drift 1 than the short-time pressure equalization strategy (6.84% vs. 16.92%, p < 0.05). However, no statistical differences were found in the incidence of drift 2 between the prolonged and short-time pressure equalization strategies (4.27% vs. 7.69%, p = 0.34). In the multivariable regression model, only the prolonged pressure equalization strategy predicted a lower incidence of pressure drift 1. In conclusion, the prolonged pressure equalization strategy was associated with a lower incidence of significant pressure drifting with more stringent thresholds than the short-time pressure equalization strategy.

原文英語
文章編號11504
期刊Scientific reports
14
發行號1
DOIs
出版狀態已發佈 - 2024 12月

ASJC Scopus subject areas

  • 多學科

指紋

深入研究「Effect of prolonged pressure equalization on final drifting during pressure wire studies」主題。共同形成了獨特的指紋。

引用此