摘要
In this work, we developed immunomagnetic detection techniques for detecting human C-reactive protein (CRP). To enhance the technique's sensitivity to human CRP, biofunctionalized magnetic nanoparticles were used as markers, and a superconductive quantum interference device gradiometer system was adapted to measure the saturated magnetization of magnetically labeled CRP-anti-CRP immune complexes. Sensitivity for human CRP was found to be as high as 1 ng in 0.1 ml, i.e., 10 ng/ml in concentration. This sensitivity is much higher than that of conventional enzyme linked immunosorbent assay by one order of magnitude.
原文 | 英語 |
---|---|
文章編號 | 252506 |
期刊 | Applied Physics Letters |
卷 | 88 |
發行號 | 25 |
DOIs | |
出版狀態 | 已發佈 - 2006 6月 19 |
ASJC Scopus subject areas
- 物理與天文學(雜項)