An implantable integrated SiGe FM transmitter for HRV biotelemetry

Fang Ren Liao*, Chi An Chen, Shey Shi Lu, Nan Fu Chiu, Chii Wann Lin, Jen Yu Li, Chia Nan Chien, Fu Shan Jaw, Jiun Min Wang, Lung Jieh Yang, Tzu Chien Hsiao, Chih Kung Lee

*此作品的通信作者

研究成果: 書貢獻/報告類型會議論文篇章

4 引文 斯高帕斯(Scopus)

摘要

This paper presents an implantable integrated FM transmitter using 0.35um SiGe process for biotelemetry. The architecture of FM transmitter is based on the Colpitts oscillator, which is followed by a buffer amplifier to isolate the influence of output on the resonant frequency of the oscillator. The chip area of the fabricated FM transmitter is only 0.26 mm 2. Even without an intentional antenna, the experimental results show that electrocardiograms (ECGs) of a rat using the FM transmitter can be successfully transmitted to a personal computer, where the heart rate variability (HRV) is analyzed.

原文英語
主出版物標題2004 IEEE International Workshop on Biomedical Circuits and Systems
頁面S1.8-9-S1.8-11
出版狀態已發佈 - 2004
對外發佈
事件2004 IEEE International Workshop on Biomedical Circuits and Systems - Singapore, 新加坡
持續時間: 2004 12月 12004 12月 3

出版系列

名字2004 IEEE International Workshop on Biomedical Circuits and Systems

其他

其他2004 IEEE International Workshop on Biomedical Circuits and Systems
國家/地區新加坡
城市Singapore
期間2004/12/012004/12/03

ASJC Scopus subject areas

  • 一般工程

指紋

深入研究「An implantable integrated SiGe FM transmitter for HRV biotelemetry」主題。共同形成了獨特的指紋。

引用此