A Temperature-Stable Cryo-System for High-Temperature Superconducting MR In-Vivo Imaging

In Tsang Lin, Hong Chang Yang*, Jyh Horng Chen

*此作品的通信作者

研究成果: 雜誌貢獻期刊論文同行評審

4 引文 斯高帕斯(Scopus)

摘要

To perform a rat experiment using a high-temperature superconducting (HTS) surface resonator, a cryostat is essential to maintain the rat's temperature. In this work, a compact temperature-stable HTS cryo-system, keeping animal rectal temperature at 37.4°C for more than 3 hours, was successfully developed. With this HTS cryo-system, a 40-mm-diameter Bi2Sr2Ca2Cu3Ox (Bi-2223) surface resonator at 77 K was demonstrated in a 3-Tesla MRI system. The proton resonant frequency (PRF) method was employed to monitor the rat's temperature. Moreover, the capacity of MR thermometry in the HTS experiments was evaluated by correlating with data from independent fiber-optic sensor temperature measurements. The PRF thermal coefficient was derived as 0.03 rad/°C and the temperature-monitoring architecture can be implemented to upgrade the quality and safety in HTS experiments. The signal-to-noise ratio (SNR) of the HTS surface resonator at 77 K was higher than that of a professionally made copper surface resonator at 300 K, which has the same geometry, by a 3.79-fold SNR gain. Furthermore, the temperature-stable HTS cryo-system we developed can obtain stable SNR gain in every scan. A temperature-stable HTS cryo-system with an external air-blowing circulation system is demonstrated.

原文英語
文章編號e61958
期刊PloS one
8
發行號4
DOIs
出版狀態已發佈 - 2013 4月 18

ASJC Scopus subject areas

  • 一般生物化學,遺傳學和分子生物學
  • 一般農業與生物科學
  • 多學科

指紋

深入研究「A Temperature-Stable Cryo-System for High-Temperature Superconducting MR In-Vivo Imaging」主題。共同形成了獨特的指紋。

引用此