摘要
In this paper, an instrumentation of the Earth's field nuclear magnetic resonance (EFNMR) inside a laboratory is presented. A lock-in analysis (LIA) technique was proposed to enhance the signal-to-noise ratio (SNR). A SNR of 137.8 was achieved in a single measurement for 9 ml tap water, and the LIA technique significantly enhanced the SNR to 188 after a 10-average in a noisy laboratory environment. The proton-phosphorus coupling in trimethyl phosphate ((CH3O)3PO) with J-coupling J [H,F] = (10.99±0.013) Hz has been demonstrated. The LIA technique improves the SNR, and a 2.6-fold improvement in SNR over that of the frequency-adjusted averaging is achieved. To reduce the noise in EFNMR, it was suggested that the LIA technique and the first order gradient shim be used to achieve a subhertz linewidth.
原文 | 英語 |
---|---|
文章編號 | 104104 |
期刊 | Review of Scientific Instruments |
卷 | 81 |
發行號 | 10 |
DOIs | |
出版狀態 | 已發佈 - 2010 10月 |
ASJC Scopus subject areas
- 儀器