Soft x-ray spectroscopy beamline 6 m high energy spherical grating monochromator at SRRC: Optical design and first performance tests

S. C. Chung, C. I. Chen, P. C. Tseng, H. F. Lin, T. E. Dann, Y. F. Song, L. R. Huang, C. C. Chen, J. M. Chuang, K. L. Tsang, C. N. Chang

Research output: Contribution to journalArticle

23 Citations (Scopus)

Abstract

The layout and first data on the performance of the soft x-ray bending magnet beamline 6 m high energy spherical grating monochromator (6m-HSGM) at SRRC are described. The 6m-HSGM beamline, which covers photon energies from 110 to 1500 eV, is based on the Dragon concept with spherical gratings, and a movable exit slit. During the first performance tests, core-excitation thresholds of Ar and Ne, and K thresholds of N and O for gas phase CO, N 2, and O2 have been recorded with resolutions which rank among the best reached so far in the soft x-ray region.

Original languageEnglish
Pages (from-to)1655-1657
Number of pages3
JournalReview of Scientific Instruments
Volume66
Issue number2
DOIs
Publication statusPublished - 1995 Dec 1

    Fingerprint

ASJC Scopus subject areas

  • Instrumentation

Cite this

Chung, S. C., Chen, C. I., Tseng, P. C., Lin, H. F., Dann, T. E., Song, Y. F., Huang, L. R., Chen, C. C., Chuang, J. M., Tsang, K. L., & Chang, C. N. (1995). Soft x-ray spectroscopy beamline 6 m high energy spherical grating monochromator at SRRC: Optical design and first performance tests. Review of Scientific Instruments, 66(2), 1655-1657. https://doi.org/10.1063/1.1145873