Simultaneous identification of the low-field induced tiny variation of complex refractive index for anisotropic and opaque magnetic-fluid thin film by a stable heterodyne mach-zehnder interferometer

Chin Yih Hong, Jen Jie Chieh, Shieh Yueh Yang, Hong Chang Yang, Herng Er Horng

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

We use a heterodyne Mach-Zehnder interferometer to simultaneously and simply measure the complex refractive index by only normal incidence on the specimen, instead of using a complicated measurement procedure or instrument that only measures the real or imaginary part of the complex refractive index. To study the tiny variation of the complex refractive index, the small complex refractive-index variation of a rare-concentration magnetic-fluid thin film, due to a weak field of less than 200 Oe, was processed by this interferometer. We also present the wavelength trend of the complex refractive index of magnetic fluids to verify the appearance of the slight change in a small wavelength range.

Original languageEnglish
Pages (from-to)5604-5611
Number of pages8
JournalApplied Optics
Volume48
Issue number29
DOIs
Publication statusPublished - 2009 Oct 10

Fingerprint

Mach-Zehnder interferometers
refractivity
fluids
thin films
wavelengths
interferometers
incidence
trends

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Engineering (miscellaneous)
  • Electrical and Electronic Engineering

Cite this

@article{0b2cd0bb5281497d800f62e7ecd464c1,
title = "Simultaneous identification of the low-field induced tiny variation of complex refractive index for anisotropic and opaque magnetic-fluid thin film by a stable heterodyne mach-zehnder interferometer",
abstract = "We use a heterodyne Mach-Zehnder interferometer to simultaneously and simply measure the complex refractive index by only normal incidence on the specimen, instead of using a complicated measurement procedure or instrument that only measures the real or imaginary part of the complex refractive index. To study the tiny variation of the complex refractive index, the small complex refractive-index variation of a rare-concentration magnetic-fluid thin film, due to a weak field of less than 200 Oe, was processed by this interferometer. We also present the wavelength trend of the complex refractive index of magnetic fluids to verify the appearance of the slight change in a small wavelength range.",
author = "Hong, {Chin Yih} and Chieh, {Jen Jie} and Yang, {Shieh Yueh} and Yang, {Hong Chang} and Horng, {Herng Er}",
year = "2009",
month = "10",
day = "10",
doi = "10.1364/AO.48.005604",
language = "English",
volume = "48",
pages = "5604--5611",
journal = "Applied Optics",
issn = "0003-6935",
publisher = "The Optical Society",
number = "29",

}

TY - JOUR

T1 - Simultaneous identification of the low-field induced tiny variation of complex refractive index for anisotropic and opaque magnetic-fluid thin film by a stable heterodyne mach-zehnder interferometer

AU - Hong, Chin Yih

AU - Chieh, Jen Jie

AU - Yang, Shieh Yueh

AU - Yang, Hong Chang

AU - Horng, Herng Er

PY - 2009/10/10

Y1 - 2009/10/10

N2 - We use a heterodyne Mach-Zehnder interferometer to simultaneously and simply measure the complex refractive index by only normal incidence on the specimen, instead of using a complicated measurement procedure or instrument that only measures the real or imaginary part of the complex refractive index. To study the tiny variation of the complex refractive index, the small complex refractive-index variation of a rare-concentration magnetic-fluid thin film, due to a weak field of less than 200 Oe, was processed by this interferometer. We also present the wavelength trend of the complex refractive index of magnetic fluids to verify the appearance of the slight change in a small wavelength range.

AB - We use a heterodyne Mach-Zehnder interferometer to simultaneously and simply measure the complex refractive index by only normal incidence on the specimen, instead of using a complicated measurement procedure or instrument that only measures the real or imaginary part of the complex refractive index. To study the tiny variation of the complex refractive index, the small complex refractive-index variation of a rare-concentration magnetic-fluid thin film, due to a weak field of less than 200 Oe, was processed by this interferometer. We also present the wavelength trend of the complex refractive index of magnetic fluids to verify the appearance of the slight change in a small wavelength range.

UR - http://www.scopus.com/inward/record.url?scp=70350326128&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=70350326128&partnerID=8YFLogxK

U2 - 10.1364/AO.48.005604

DO - 10.1364/AO.48.005604

M3 - Article

C2 - 19823245

AN - SCOPUS:70350326128

VL - 48

SP - 5604

EP - 5611

JO - Applied Optics

JF - Applied Optics

SN - 0003-6935

IS - 29

ER -