Monochromator alignment technique

King Long Tsang, Ping Chung Tseng, Ching Iue Chen, Yen Fang Song, Shinchun Chung, Tang Eh Dann, Hsueh Feng Lin, Liang Ren Huang, Chu Nan Chang

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

A simple and high precision technique is developed for the monochromator alignment. The alignment setup is composed of two He/Ne lasers, a penta-prism, a flat mirror, and other common alignment aids. The alignment principle and the setup is described. An example of how to align Synchrotron Radiation Research Center 1-m-Seya monochromator is presented as well.

Original languageEnglish
Pages (from-to)1293-1295
Number of pages3
JournalReview of Scientific Instruments
Volume63
Issue number1
DOIs
Publication statusPublished - 1992 Dec 1

    Fingerprint

ASJC Scopus subject areas

  • Instrumentation

Cite this

Tsang, K. L., Tseng, P. C., Chen, C. I., Song, Y. F., Chung, S., Dann, T. E., Lin, H. F., Huang, L. R., & Chang, C. N. (1992). Monochromator alignment technique. Review of Scientific Instruments, 63(1), 1293-1295. https://doi.org/10.1063/1.1143103