Abstract
We applied prepolarization field and high-Tc superconducting quantum interference device (SQUID) detector to enhance nuclear magnetic resonance signal in a microtesla magnetic field. The minimum measuring magnetic field is 8.9 μT at which the proton resonance frequency is 380 Hz. The specificity instrumentation and the difficulty of using a high-Tc SQUID with prepolarization field were investigated. We applied gradient field to perform one-dimensional proton imaging in a microtesla magnetic field. Additionally, low field high-Tc SQUID-based NMR systems are promising in biomagnetic research due to its use, for example, in imaging with hyperpolarized noble gas.
Original language | English |
---|---|
Article number | 252505 |
Journal | Applied Physics Letters |
Volume | 88 |
Issue number | 25 |
DOIs | |
Publication status | Published - 2006 Jun 19 |
ASJC Scopus subject areas
- Physics and Astronomy (miscellaneous)