Abstract
A simple, inexpensive and reliable method for the simultaneous, routine analysis of riboflavin in urine by capillary electrophoresis-light emitting diode (LED)-induced fluorescence detection is described. Using a blue LED as the light source, the detection limit of riboflavin was determined to be 0.48 μg/ml and was improved to 20 ng/ml when a stacking technique was applied. In the analysis of an actual sample, various concentrations of riboflavin were distributed in the urine samples over a period of 9 h after the ingestion of a vitamin B2 tablet.
| Original language | English |
|---|---|
| Pages (from-to) | 39-46 |
| Number of pages | 8 |
| Journal | Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences |
| Volume | 785 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 2003 Feb 25 |
Keywords
- Blue light emitting diode
- Riboflavin
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Clinical Biochemistry
- Cell Biology
Fingerprint
Dive into the research topics of 'Determination of riboflavin in urine by capillary electrophoresis-blue light emitting diode-induced fluorescence detection combined with a stacking technique'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS